Warnings, Precautions & Limitations
• Youmaygettestresultshigherthanyouractualglucoselevel
if taking PAM (1-Methylpyridine-6-carbaldehyde oxime). Taking
unnecessary measures to lower your blood glucose level may induce
severe low blood glucose symptoms such as coma.
•Nevermakesignicantchangestoyourdiabetescontrolprogram
or ignore physical symptoms without consulting with your healthcare
professional.
•Patientsundergoingoxygentherapymayyieldfalseresults.
• Heparinsamplesofbloodmaybeused.Donotuseother
preservatives.
•Donotgraspthemeterneartheteststripport.Abuilt-in
thermosensor used to minimize measurement errors is installed just
inside the port. If you hold this area, it will be warmed by your fingers,
which will adversely affect test results.
•Severedehydration(excessivewaterloss)maycausefalselow
results. If you believe you are suffering from dehydration, consult your
healthcare professional right away.